Summary

International Symposium on Electromagnetic Compatibility

2014

Session Number:14P2-H

Session:

Number:14P2-H6

Safety Assessment of AIMD under MRI Exposure: Tier-3 vs. Tier-4 Evaluation of Local RF-induced Heating

Eugenia Cabot,  Earl Zastrow,  Niels Kuster,  

pp.-

Publication Date:2014/05/12

Online ISSN:2188-5079

DOI:10.34385/proc.18.14P2-H6

PDF download (635.4KB)

Summary:
Patients with active implantable medical devices (AIMDs) are generally excluded from magnetic resonance (MR) diagnostics because interference of the AIMD with MR-induced radiofrequency (RF) fields can lead to hazardous localized heating in the surrounding tissues. In this work, we report the safety assessment of a generic device performed according to tier 3 and 4 guidelines in Technical Specification 10974. For tier 3, local tissue heating was obtained by applying RF fields, obtained numerically in vivo in human models exposed to MRI coils, to a transfer function of the device. For tier 4, a validated model of the generic device was implanted in anatomical human models exposed to the RF of MRI coils, and full-wave computational electromagnetic simulations were performed. A comparison of the two methods is made.