Presentation | 2014-12-13 Examination of an indicated and an estimate algorithm calculated blood pressure value in the oscillometric sphygmomanometer Yasushi Makka, Satoshi Kuroda, Toshitsugu Sugawara, Takeshi Tsuruga, Kazuyuki Kimura, Hisae O. Shimizu, |
---|---|
PDF Download Page | PDF download Page Link |
Abstract(in Japanese) | (See Japanese page) |
Abstract(in English) | Blood pressure measurement is important to know the lifestyle-related diseases. In the "hypertension guidelines 2014", there is a "case of blood pressure in the hospital and at home is different priority to blood pressure in the home.", Came to the home in the blood pressure is important. Sphygmomanometer are both used in the medical facilities and the home is the oscillometric sphygmomanometer. Research on the accuracy of blood pressure values oscillometric blood pressure monitor to display less. In this study, we clarify the difference of the algorithm from publications that are patented published. Furthermore, using a simulator that can output a constant pattern signal, and examined the pressure pulse wave of differences in oscillometric sphygmomanometer. As a result, from the publications are published patents, there is a difference in blood pressure value displayed between companies. Although each company are able to detect along the algorithm, although there is a tolerance range it was confirmed that there is a variation in the measured value by the decision algorithm is different. |
Keyword(in Japanese) | (See Japanese page) |
Keyword(in English) | oscillometric sphygmomanometer / algorithm / Display value variation |
Paper # | MBE2014-90 |
Date of Issue |
Conference Information | |
Committee | MBE |
---|---|
Conference Date | 2014/12/6(1days) |
Place (in Japanese) | (See Japanese page) |
Place (in English) | |
Topics (in Japanese) | (See Japanese page) |
Topics (in English) | |
Chair | |
Vice Chair | |
Secretary | |
Assistant |
Paper Information | |
Registration To | ME and Bio Cybernetics (MBE) |
---|---|
Language | JPN |
Title (in Japanese) | (See Japanese page) |
Sub Title (in Japanese) | (See Japanese page) |
Title (in English) | Examination of an indicated and an estimate algorithm calculated blood pressure value in the oscillometric sphygmomanometer |
Sub Title (in English) | |
Keyword(1) | oscillometric sphygmomanometer |
Keyword(2) | algorithm |
Keyword(3) | Display value variation |
1st Author's Name | Yasushi Makka |
1st Author's Affiliation | Division of Biomedical Engineering, Graduate School of Hokkaido University of Science() |
2nd Author's Name | Satoshi Kuroda |
2nd Author's Affiliation | Division of Biomedical Engineering, Graduate School of Hokkaido University of Science |
3rd Author's Name | Toshitsugu Sugawara |
3rd Author's Affiliation | Division of Biomedical Engineering, Graduate School of Hokkaido University of Science |
4th Author's Name | Takeshi Tsuruga |
4th Author's Affiliation | Division of Biomedical Engineering, Graduate School of Hokkaido University of Science |
5th Author's Name | Kazuyuki Kimura |
5th Author's Affiliation | Division of Biomedical Engineering, Graduate School of Hokkaido University of Science |
6th Author's Name | Hisae O. Shimizu |
6th Author's Affiliation | Division of Biomedical Engineering, Graduate School of Hokkaido University of Science |
Date | 2014-12-13 |
Paper # | MBE2014-90 |
Volume (vol) | vol.114 |
Number (no) | 361 |
Page | pp.pp.- |
#Pages | 6 |
Date of Issue |