Presentation | 2001/5/24 Doubts and Anxiety with Grounding for Safety of Medical Electrical Equipment Norio Murasaki, |
---|---|
PDF Download Page | PDF download Page Link |
Abstract(in Japanese) | (See Japanese page) |
Abstract(in English) | In order to prevent microshock of patient in medical facilities, equepotential patient reference system set potential level under 10mV. 10mV was introduced by a assumption : Electric shock is a function of leak current or contact potential difference of some metals. Some nonmetalic materiales has powerful e.m.f. It is need to keep potential of everythings in medecal facilities at the same potential of patient heart. |
Keyword(in Japanese) | (See Japanese page) |
Keyword(in English) | Medical Electrical Equipment / Equipotential Patient Reference System / Electrical Shock / Microshock / Voltage Compensate Circuit / Safety |
Paper # | SSS2001-1 |
Date of Issue |
Conference Information | |
Committee | SSS |
---|---|
Conference Date | 2001/5/24(1days) |
Place (in Japanese) | (See Japanese page) |
Place (in English) | |
Topics (in Japanese) | (See Japanese page) |
Topics (in English) | |
Chair | |
Vice Chair | |
Secretary | |
Assistant |
Paper Information | |
Registration To | Safety (SSS) |
---|---|
Language | JPN |
Title (in Japanese) | (See Japanese page) |
Sub Title (in Japanese) | (See Japanese page) |
Title (in English) | Doubts and Anxiety with Grounding for Safety of Medical Electrical Equipment |
Sub Title (in English) | |
Keyword(1) | Medical Electrical Equipment |
Keyword(2) | Equipotential Patient Reference System |
Keyword(3) | Electrical Shock |
Keyword(4) | Microshock |
Keyword(5) | Voltage Compensate Circuit |
Keyword(6) | Safety |
1st Author's Name | Norio Murasaki |
1st Author's Affiliation | Tokyo University of Agriculture and Technology() |
Date | 2001/5/24 |
Paper # | SSS2001-1 |
Volume (vol) | vol.101 |
Number (no) | 102 |
Page | pp.pp.- |
#Pages | 6 |
Date of Issue |