Presentation | 2015-11-13 Verification of high definition of low field MRI image using NHA Masaya Hasegawa, Shigeki Hirobayashi, Tadanobu Misawa, Koichi Mori, Kyo Noguchi, |
---|---|
PDF Download Page | PDF download Page Link |
Abstract(in Japanese) | (See Japanese page) |
Abstract(in English) | In late years the MRI that is a tomogram technology using a magnetic resonance phenomenon attracts its attention. MRI acquires biological information as a complex signal, and the image reconstruction by the FFT. Therefore, a point image function after the FFT becomes clear and improvement of the space resolution if a period of enough length enters the measurement signal. In other words, an high magnetic field is essential to achieve high resolution in FFT. However, high magnetic fields of which leads to maintenance costs increase and upsizing of the device. In this study, we extend the NHA that is high accuracy signal analysis method to the complex signal. Then, we show the possibility of providing high definition image at low field MRI by the computer simulation that assumed ideal environment. |
Keyword(in Japanese) | (See Japanese page) |
Keyword(in English) | Magnetic Resonance Imaging / Non-harmonic Analysis / Low magnetic field measurement |
Paper # | IE2015-81 |
Date of Issue | 2015-11-05 (IE) |
Conference Information | |
Committee | IE / ITE-ME / ITE-AIT |
---|---|
Conference Date | 2015/11/12(2days) |
Place (in Japanese) | (See Japanese page) |
Place (in English) | |
Topics (in Japanese) | (See Japanese page) |
Topics (in English) | |
Chair | Seishi Takamura(NTT) / Kazuhito Murakami(Aichi Prefectural Univ.) / Tokiichiro Takahashi(TDU) |
Vice Chair | Takayuki Hamamoto(Tokyo Univ. of Science) / Atsuro Ichigaya(NHK) / Miki Haseyama(Hokkaido Univ.) |
Secretary | Takayuki Hamamoto(NTT) / Atsuro Ichigaya(Chiba Inst. of Tech.) / Miki Haseyama(NTT) / (Niigata Univ.) |
Assistant | Keita Takahashi(Nagoya Univ.) / Kei Kawamura(KDDI R&D Labs.) |
Paper Information | |
Registration To | Technical Committee on Image Engineering / Technical Group on Media Engineering / Technical Group on Artistic Image Technology |
---|---|
Language | JPN |
Title (in Japanese) | (See Japanese page) |
Sub Title (in Japanese) | (See Japanese page) |
Title (in English) | Verification of high definition of low field MRI image using NHA |
Sub Title (in English) | |
Keyword(1) | Magnetic Resonance Imaging |
Keyword(2) | Non-harmonic Analysis |
Keyword(3) | Low magnetic field measurement |
1st Author's Name | Masaya Hasegawa |
1st Author's Affiliation | University of Toyama(Toyama Univ.) |
2nd Author's Name | Shigeki Hirobayashi |
2nd Author's Affiliation | University of Toyama(Toyama Univ.) |
3rd Author's Name | Tadanobu Misawa |
3rd Author's Affiliation | University of Toyama(Toyama Univ.) |
4th Author's Name | Koichi Mori |
4th Author's Affiliation | Toyama University Hospital(Toyama Univ.Hospital) |
5th Author's Name | Kyo Noguchi |
5th Author's Affiliation | University of Toyama(Toyama Univ.) |
Date | 2015-11-13 |
Paper # | IE2015-81 |
Volume (vol) | vol.115 |
Number (no) | IE-308 |
Page | pp.pp.41-45(IE), |
#Pages | 5 |
Date of Issue | 2015-11-05 (IE) |