Presentation 2023-06-16
Investigation of contamination in Heater-cooler unit water systems for cardiopulmonary bypass.
Koya Sakaitani, Minoru Nakamura, Daisuke Furuya, Toshitsugu Sugawara, Indou Tomokazu,
PDF Download Page PDF download Page Link
Abstract(in Japanese) (See Japanese page)
Abstract(in English) In 2015, Mycobacterium chimaera infection was reported in a patient who had undergone open heart surgery, and it was revealed that the cause was bacterial contamination of circulating water in the Heater-Cooler Unit (HCU) for cardiopulmonary bypass. This led to the recommendation of hygiene management for devices with water circuits in operating rooms. However, even now, there are no guidelines, and the method is left to each hospital. Therefore, for the purpose of selecting appropriate hygiene management, we first attempted to identify and classify bacteria in circulating water in HCUs at three medical institutions to determine the bacterial contamination status of devices used in clinical settings. As a result, glucose non-fermentative Gram-negative rods were commonly detected in all hospitals. Therefore, it was suggested that HCUs may be a potential source of infection when sanitation and operation of the HCUs are inappropriate.
Keyword(in Japanese) (See Japanese page)
Keyword(in English) Heater-Cooler Unit / Extracorporeal circulation / Hygiene management / Surgical site infection
Paper # MBE2023-13
Date of Issue 2023-06-09 (MBE)

Conference Information
Committee MBE / IEE-MBE
Conference Date 2023/6/16(1days)
Place (in Japanese) (See Japanese page)
Place (in English) Hokkaido University
Topics (in Japanese) (See Japanese page)
Topics (in English) ME, etc.
Chair Hisashi Yoshida(Kinki Univ.)
Vice Chair Akinori Ueno(Tokyo Denki Univ.)
Secretary Akinori Ueno(Tohoku Inst. of Tech.) / (Univ. of Tsukuba)
Assistant Miki Kaneko(Osaka Univ.) / Akihiko Tsukahara(Tokyo Denki Univ.)

Paper Information
Registration To Technical Committee on ME and Bio Cybernetics / The Technical Committee on Medical and Biological Engineering
Language JPN
Title (in Japanese) (See Japanese page)
Sub Title (in Japanese) (See Japanese page)
Title (in English) Investigation of contamination in Heater-cooler unit water systems for cardiopulmonary bypass.
Sub Title (in English) Investigation report on clinical devvices
Keyword(1) Heater-Cooler Unit
Keyword(2) Extracorporeal circulation
Keyword(3) Hygiene management
Keyword(4) Surgical site infection
1st Author's Name Koya Sakaitani
1st Author's Affiliation Hokkaido University of Science(HUS)
2nd Author's Name Minoru Nakamura
2nd Author's Affiliation Hokkaido University of Science(HUS)
3rd Author's Name Daisuke Furuya
3rd Author's Affiliation Hokkaido University of Science(HUS)
4th Author's Name Toshitsugu Sugawara
4th Author's Affiliation Hokkaido University of Science(HUS)
5th Author's Name Indou Tomokazu
5th Author's Affiliation Hokkaido University of Science(HUS)
Date 2023-06-16
Paper # MBE2023-13
Volume (vol) vol.123
Number (no) MBE-82
Page pp.pp.22-23(MBE),
#Pages 2
Date of Issue 2023-06-09 (MBE)