Presentation | 2020-01-30 Flow-velocity measurement accuracy of three-dimensional cine phase contrast magnetic resonance imaging using a parallel imaging technique Atsushi Fukuyama, Katsuhiro Inoue, Yutaka kato, |
---|---|
PDF Download Page | PDF download Page Link |
Abstract(in Japanese) | (See Japanese page) |
Abstract(in English) | The purpose of this study is to validate the accuracy of measuring flow velocity with parallel imaging technique, which is effective in reducing acquisition time. We performed 3D cine phase-contrast MR imaging with prospective pseudo ECG gating using a 3-Tesla MR imaging scanner with 12ch head matrix and neck matrix receiving coils. The parallel imaging technique (PAT) factors was changed from 2 to 4. As the PAT factor increased, the spatially-averaged velocity error rate increased, but never exceeded 10%. In the PAT factors used in this study, no significant measurement error was observed, so this technique can contribute to save the acquisition time. |
Keyword(in Japanese) | (See Japanese page) |
Keyword(in English) | 3D cine phase contrast magnetic resonance imaging / hemodynamic analysis / parallel imaging technique / the spatially-averaged velocity / the maximum velocity |
Paper # | MI2019-105 |
Date of Issue | 2020-01-22 (MI) |
Conference Information | |
Committee | MI |
---|---|
Conference Date | 2020/1/29(2days) |
Place (in Japanese) | (See Japanese page) |
Place (in English) | OKINAWAKEN SEINENKAIKAN |
Topics (in Japanese) | (See Japanese page) |
Topics (in English) | Medical Image Engineering, Analysis, Recognition, etc. |
Chair | Yoshiki Kawata(Tokushima Univ.) |
Vice Chair | Takayuki Kitasaka(Aichi Inst. of Tech.) / Hidekata Hontani(Nagoya Inst. of Tech.) |
Secretary | Takayuki Kitasaka(Yamaguchi Univ.) / Hidekata Hontani(Univ. of Hyogo) |
Assistant | Hotaka Takizawa(Tsukuba Univ.) / Yoshito Otake(NAIST) |
Paper Information | |
Registration To | Technical Committee on Medical Imaging |
---|---|
Language | JPN |
Title (in Japanese) | (See Japanese page) |
Sub Title (in Japanese) | (See Japanese page) |
Title (in English) | Flow-velocity measurement accuracy of three-dimensional cine phase contrast magnetic resonance imaging using a parallel imaging technique |
Sub Title (in English) | Verification using the flow phantom capable of measuring flow velocity in three directions |
Keyword(1) | 3D cine phase contrast magnetic resonance imaging |
Keyword(2) | hemodynamic analysis |
Keyword(3) | parallel imaging technique |
Keyword(4) | the spatially-averaged velocity |
Keyword(5) | the maximum velocity |
1st Author's Name | Atsushi Fukuyama |
1st Author's Affiliation | Nagoya University(Nagoya Univ.) |
2nd Author's Name | Katsuhiro Inoue |
2nd Author's Affiliation | Nagoya University Hospital(Mie Univ.) |
3rd Author's Name | Yutaka kato |
3rd Author's Affiliation | Nagoya University Hospital(Nagoya Univ.) |
Date | 2020-01-30 |
Paper # | MI2019-105 |
Volume (vol) | vol.119 |
Number (no) | MI-399 |
Page | pp.pp.177-180(MI), |
#Pages | 4 |
Date of Issue | 2020-01-22 (MI) |