Paper Abstract and Keywords |
Presentation |
2017-03-14 13:10
Microcirculatory velocity estimation with radon transformation for confocal laser scanning microscopic image Kazuhiro Nakamura, Mutoh Tatsushi, Kazumasu Sasaki (Akita Noken/Tohoku Univ.), Yasuyuki Taki (Tohoku Univ.), Tatsuya Ishikawa (Akita Noken) MBE2016-96 |
Abstract |
(in Japanese) |
(See Japanese page) |
(in English) |
Radon transformation for line-scan image of the confocal laser-scanning microscopy are a means to observe the velocity of red blood cells in the brain micro-vessels. Preprocessing for the high contrast image and edge extraction was applied for the low signal noise ratio image. Two hundred and ninety images from twenty-two animals were investigated. We can estimate the velocity from all the images even some threshold value and extracted region must be manually setting. Even it is not completely auto-determination, we can apply the methods for objective determination in the investigation of our animal studies. |
Keyword |
(in Japanese) |
(See Japanese page) |
(in English) |
Microcirculatory velocity / Confocal laser scanning microscope / Radon trnsform / / / / / |
Reference Info. |
IEICE Tech. Rep., vol. 116, no. 520, MBE2016-96, pp. 71-74, March 2017. |
Paper # |
MBE2016-96 |
Date of Issue |
2017-03-06 (MBE) |
ISSN |
Print edition: ISSN 0913-5685 Online edition: ISSN 2432-6380 |
Copyright and reproduction |
All rights are reserved and no part of this publication may be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopy, recording, or any information storage and retrieval system, without permission in writing from the publisher. Notwithstanding, instructors are permitted to photocopy isolated articles for noncommercial classroom use without fee. (License No.: 10GA0019/12GB0052/13GB0056/17GB0034/18GB0034) |